Rapid temperature change (high humidity) tester
5.0
Read 219 reviewsI. Relevant standard test requirements1. temperature ictus1.1The test object is battery pack or system1.2The lithium ion battery pack or system is placed in an alternating temperature environment ℃(-40±2)(85) and the conversion time of the two extreme temperatures is less than 30 min.For each extr
17 people are viewing this right now
I. Relevant standard test requirements
1. temperature ictus
1.1The test object is battery pack or system
1.2The lithium ion battery pack or system is placed in an alternating temperature environment ℃(-40±2)(85) and the conversion time of the two extreme temperatures is less than 30 min.
For each extreme temperature environment, the test object h, kept 8 cycles 5 times
1.3After the test ,2 h should be observed at room temperature
2.damp-heat cycle
2.1The test object is battery pack or system
II. Main technical parameters
Product model | GN-KS5-12m³-C |
Inner dimension | W2000 X D3000 X H2000 MM |
Outline dimension | W2400 X D4500 X H2940 MM |
Temperature range | -50℃~+150℃( adjustable) |
Humidity range | 20%RH-98%RH( limiting temperature ) |
Temperature fluctuation | ±0.5℃ |
Telative humidity fluctuation | ±2.5%RH |
Temperature departure | ±2.0℃ |
Humidity deviation | A)>75%RH:≤﹢2,-3%RH;B)<75%RH:≤±5%RH |
Heating rate | 5℃/MIN(-40℃~+85℃,Non-linear load 800KG) |
Cooling rate | 5℃/MIN(+85℃~ -40℃,Non-linear load 800KG) |
Lift temperature overshoot | <2℃ |
Gross weight of equipment | About 5 tons |
Device power supply (DPS) | AC 380V 127KW |